TY - JOUR ID - 11881 TI - A local study on diagnostic reference levels estimation for children computed tomography JO - Iranian Journal of Medical Physics JA - IJMP LA - en SN - AU - Afzalipour, Reza AU - Abdollahi, Hamid AU - Jafari, Salman AU - Mahdavi, Seied Rabi AD - Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran AD - Department of Radiology Technology, School of Paramedicine, Hamadan University of Medical Sciences, Hamadan, Iran Y1 - 2018 PY - 2018 VL - 15 IS - Special Issue-12th. Iranian Congress of Medical Physics SP - 26 EP - 26 KW - DRL KW - CT Scan Tehran KW - Dose Estimation DO - 10.22038/ijmp.2018.11881 N2 - Introduction: The Diagnostic reference levels (DRLs) play a critical role in the optimization of radiation dose especially, in some conditions like pediatrics. They are useful indicators by which the radiologists can be aware of delivered excess radiation doses to the patients, and take corrective actions if necessary. In order to meet some requirements for establishing the national computed tomography DRLs tables, much studies are needed all around the country. Materials and Methods: All active computed tomography (CT) scanners in public and private centers in Tehran were identified and checked for quality assurance and control certification. Eleven centers were chosen to be studied according to CT examination frequencies. Weighted CT dose index (CTDIw) and dose length product (DLP) for head, sinus, chest and abdomen/pelvis scans of children were obtained from scanner’s operator consoles and classified into four groups based on their ages (A; <1 year, B; 1-5 years, C; 5-10 years and D; 10-15 years). The 3rd quartiles of CTDIw were considered as DRLs and compared with the reported European Union (EU) and United Kingdom (UK) ones. Results: DRLs for head, sinus, chest and abdomen/pelvis scans were found to be 86.76, 31.33, 6.33, 7.65 mGy; 43.38, 31.33, 6.33, 7.65 mGy; 43, 31.33, 6.33, 7.65 mGy and 44.53, 31.33, 6.33, 7.65 mGy in the four groups (A-D) respectively. They are lower than the reported DRLs in EU and UK. Conclusion: There are variations in the radiation dose between the CT centers and identical scanners indicating the necessity for dose optimization. The data reported in this study can be remarkably useful in this concern. UR - https://ijmp.mums.ac.ir/article_11881.html L1 - ER -