Dose Distribution Evaluation and Independent Quality Check of Spherical INTRABEAM™ Applicators via Radiochromic EBT2 Film Measurement

Document Type : Original Paper

Authors

1 Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

2 Medical Physics Research Centere, Faculty of medicine, Mashhad University of Medical Sciences, Mashhad, Iran

3 Cancer Research Center, Mashhad University of Medical Sciences, Mashhad, Iran

Abstract

Introduction: The present study was conducted to implement a simple practical independent quality check of depth dose and isotropy of the Intrabeam™ therapeutic X-ray machine using radiochromic EBT2 film.
Material and Methods: Theindependent quality check of 1.5, 3.5, and 5-cm spherical Intrabeam™ applicators was accomplished using particular EBT2 film cutting pieces with internal rounded edges in a water phantom. Prior to this measure, the film was calibrated at three distances from the 5-cm applicator in water to clarify the effects of beam spectrum and dose rate alteration on film response. To this end, three calibration curves were plotted.
Results: The results of the one-way analysis of variance showed a critical difference between film pieces receiving equal doses at various distances (p <0.05). Therefore, depth dose curves were designed using all three calibration curves. Smaller applicators represented steeper dose fall-off, compared to the larger sizes. In this regard, 14.97%, 17.59%, and 30.92% of the relative mean doses were measured at 1 cm depth of 1.5-cm, 3.5-cm, and 5-cm applicators, respectively. A 10%/1mm gamma index was satisfied for the lateral dose evaluation of corresponding depth relative to Z-direction.
Conclusion: The approach implemented in this study could be carried out as a rapid monthly quality check method for the dose distribution evaluation oftheIntrabeam device.

Keywords

Main Subjects


  1. References

     

    1. Gunderson LL, Willett CG, Calvo FA, Harrison LB. Intraoperative irradiation: techniques and results: Springer Science & Business Media; 2011.
    2. Hensley FW. Present state and issues in IORT Physics. Radiation Oncology. 2017;12(1):37.
    3. Njeh CF, Saunders MW, Langton CM. Accelerated Partial Breast Irradiation (APBI): A review of available techniques. Radiation Oncology. 2010;5(1):90.
    4. Park CC, Yom SS, Podgorsak MB, Harris E, Price RA, Bevan A, et al. American Society for Therapeutic Radiology and Oncology (ASTRO) emerging technology committee report on electronic brachytherapy. International Journal of Radiation Oncology* Biology* Physics. 2010;76(4):963-72.
    5. Eaton DJ. Quality assurance and independent dosimetry for an intraoperative x-ray device. Med Phys. 2012;39(11):6908-20.
    6. Andres C, Del Castillo A, Tortosa R, Alonso D, Barquero R. A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT. Medical physics. 2010;37(12):6271-8.
    7. Arjomandy B, Tailor R, Anand A, Sahoo N, Gillin M, Prado K, et al. Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies. Medical physics. 2010;37(5):1942-7.
    8. Todorovic M, Fischer M, Cremers F, Thom E, Schmidt R. Evaluation of GafChromic EBT prototype B for external beam dose verification. Med Phys. 2006;33(5):1321-8.
    9. Schneider F, Polednik M, Wolff D, Steil V, Delana A, Wenz F, et al. Optimization of the gafchromic EBT protocol for IMRT QA. Zeitschrift fur medizinische Physik. 2009;19(1):29-37.
    10. Butson MJ, Peter K, Cheung T, Alnawaf H. Energy response of the new EBT2 radiochromic film to x-ray radiation. Radiation Measurements. 2010;45(7):836-9.
    11. Massillon-JL G, Chiu-Tsao S-T, Domingo-Muñoz I, Chan MF. Energy dependence of the new Gafchromic EBT3 film: dose response curves for 50 kV, 6 and 15 MV X-ray beams. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology. 2012;1(02):60.
    12. Dreindl R, Georg D, Stock M. Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films. Zeitschrift fur medizinische Physik. 2014;24(2):153-63.
    13. Carrasco MA, Perucha M, Luis FJ, Baeza M, Herrador M. A comparison between radiochromic EBT2 film model and its predecessor EBT film model. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2013;29(4):412-22.
    14. Powell J. CO2 Laser Cutting. Second Edition ed. Mordike BL, editor. London: Springer. 1998; 258.
    15. Lindsay P, Rink A, Ruschin M, Jaffray D. Investigation of energy dependence of EBT and EBT-2 gafchromic film. Med Phys. 2010;37(2):571-6.
    16. Reinhardt S, Hillbrand M, Wilkens J, Assmann W. Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. Medical physics. 2012;39(8):5257-62.
    17. Ebert MA, Asad AH, Siddiqui SA. Suitability of radiochromic films for dosimetry of low energy X‐rays. Journal of applied clinical medical physics. 2009;10(4):232-40.
    18. Ebert M, Carruthers B, Lanzon P, Haworth A, Clarke J, Caswell N, et al. Dosimetry of a low-kV intra-operative X-ray source using basic analytical beam models. Australasian Physics & Engineering Sciences in Medicine. 2002;25(3):119-23.
    19. Ramachandran P. New era of electronic brachytherapy. World journal of radiology. 2017;9(4):148.
    20. Eaton DJ, Duck S. Dosimetry measurements with an intra-operative x-ray device. Physics in medicine and biology. 2010;55(12): 359-69.
    21. Eaton DJ, Best B, Brew-Graves C, Duck S, Ghaus T, Gonzalez R, et al. In vivo dosimetry for single-fraction targeted intraoperative radiotherapy (TARGIT) for breast cancer. International Journal of Radiation Oncology• Biology• Physics. 2012;82(5):819-24.
    22. ICRP A. Early and late effects of radiation in normal tissues and organs: threshold doses for tissue reactions and other non-cancer effects of radiation in a radiation protection context. 2011.
    23. Eaton DJ, Earner B, Faulkner P, Dancer N. A national dosimetry audit of intraoperative radiotherapy. The British journal of radiology. 2013;86(1032):20130447.
    24. Hartl PM, Dobler B, Kolbl O, Treutwein M. [Practical dosimetry and constancy check at introduction of intraoperative radiotherapy with Intrabeam (Zeiss)]. Zeitschrift fur medizinische Physik. 2009;19(4):288-93.