Determination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation

Document Type : Original Paper

Authors

1 Physics,Faculty of sciences, University Mohammed V Rabat

2 Departement of Physics, Laboratory of High Energy Physics, Modelling and Simulation, Faculty of Science, Mohammed V Agdal University, Rabat, Kingdom of Morocco

3 Physics, faculty of sciences, University Mohammed V Rabt

4 Academic hospital of Udine

Abstract

Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method using the SIMIND Monte Carlo simulation code in Gadolinium-159 (Gd-159) imaging.
Material and Methods: The energy window was set for various main energy window widths (i.e., 10%, 15%, and 20%) and sub-energy window widths (i.e., 3 and 6 keV).Siemens Medical System Symbia fitted with a high-energy collimator was used with Gd-159 point source positioned at seven locations inside the cylindrical water phantom. A comparison was made between the true primary to total ratio (calculated by SIMIND) and the primary to total ratio estimated using TEW method.
Results: The findings of this study showed that 20% of the main energy windows with 3 and 6 keV sub-energy windows were optimal for the implementation of the TEW method in Gd-159.
Conclusion: According to the results, the optimal energy windows for Gd-159 scintigraphy were the sub-energy windows of 3 and 6 keV. These findings could be helpful in the quantification of Gd-159 imaging.Purpose : In radio-nuclides imaging, object scatter is one of major factors which leads to degradation of image quality. Therefore, the correction of scattered photons has a great impact to improve the image quality. The aim of this work was to determine the main and sub-energy windows for the triple energy window (TEW) scatter correction method using Monte Carlo simulation SIMIND code for Gadolinium-159 (Gd-159) imaging.
Methods : Energy window was set for various main energy window width (10,15 and 20%) and sub energy window width (3 and 6 keV). Siemens Medical System Symbia fitted with High Energy collimator (HE) was imaged with Gd-159 point source positioned at seven locations inside cylindrical water phantom. The true primary to total ratio (calculated by SIMIND) and the primary to total ratio estimated using TEW method were compared.
Results : A 20% of main energy window with 3 and 6 keV sub-energy windows were found to be optimal for implementation of the TEW method in Gd-159.
Conclusion: The obtained results provide the optimal energy window for Gd-159 scintigraphy data and will aid the quantification of Gd-159 imaging.

Keywords

Main Subjects


  1. Ogawa K, Harata Y, Ichihara T, Kubo A, Hashimoto S. Estimation of scatter component in SPECT planar image using a Monte Carlo method. Jpn J Nucl Med. 1990; 27: 467-75.
  2. Floyd CE, Jaszczak RJ, Harris CC, Coleman RE. Energy and spatial distribution of multiple order Compton scatter in SPECT: a Monte Carlo simulation. Phys Med Biol. 1984; 29:1217-30.
  3. Floyd CE, Jaszczak RJ, Coleman RE. Inverse Monte Carlo: A Unified Reconstruction Algorithm for SPECT. IEEE Trans Nucl Sci.1985; 32: 779-85.
  4. Sadremomtaz A, Telikani Z. Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study. Iran J Nucl Med. 2016; 24(2):136-43.
  5. Razavi SH, Kalantari F, Bagheri M, Namiranian N, Nafisi-Moghadam R , Mardanshahi A, et al.  Characterization of low, medium and high energy collimators for common isotopes in nuclear medicine: A Monte Carlo study. Iran J Nucl Med. 2017; 25(2):100-4.
  6. Jaszczak RJ, Greer KL, Floyd CE Jr, Harris CC, Coleman RE. Improved SPECT quantification using compensation for scattered photons. J Nucl Med.1984; 25:893–900.
  7. Ogawa K, Harata Y, Ichihara T, Kubo A, Hashimoto S. A practical method for position-dependent Compton-scatter correction in single photon emission CT. IEEE Trans. Med. Imaging. 1991; 10: 408.
  8. Lagerburg V, de Nijs R, Holm S, Svarer C. A comparison of different energy window subtraction methods to correct for scatter and down scatter in I-123 SPECT imaging. Nucl Med Commun. 2012; 33(7):708-18.
  9. Axelsson B, Msaki B, Israelsson A. Subtraction of Compton scattered photons in single-photon emission computerized tomography. J Nucl Med. 1984; 25: 490-4
  10. Hamill JJ, DeVito RP. Scatter reduction with energy weighted acquisition. IEEE Trans Nucl Sci.1989; 36: 1334-9.
  11. Koral KF, Wang X, Rogers WL, Clinthorne NH, Wang X. SPECT Compton-scattering correction by analysis of energy spectra. J Nucl Med. 1988; 29: 195-202.
  12. Gagnon D, Todd-Pokropek AE, Arsenault A, Dupras G. Introduction to holospectral imaging in nuclear medicine for scatter subtraction. IEEE Trans Med Imag. 1989; 8: 245-50.
  13. 13- Floyd CE, Jaszczak RJ, Coleman RE. Scatter detection in SPECT imaging: dependence on source depth, energy, and energy window. Phys Med Biol. 1988; 33:1075-81.
  14. Koral KF, Clinthorne NH, Rogers WL. Improving emission-computed-tomography quantification by Compton-scatter rejection through offset window. Nucl Inst and Methods in Phys Res A. 1986; 242: 610-4.
  15. Soares DCF, Menezes MA de BC, dos Santos RG, Ramaldes GA.  Gd-159: preparation and preliminary evaluation asa potential antitumoral radionuclide. J Radioanal Nucl Chem. 2010; 284: 315-20.
  16. Soares DCF, Sousa GFD, Barros ALBD  , Cardoso VN,   Oliveira MCD , Ramaldes GA.  Scintigraphic imaging and increment in mice survival using theranostic liposomes based on Gadolinium-159. J of Dru Del Science and Tech. 2015; 30:7-14.
  17. Lee YS, Kim JS, Kim KM, Lima SM , Kimb HJ. Determination of energy windows for the triple energy window scatter correction method in I-131 on a Siemens SYMBIA gamma camera: a GATE simulation study. J of Inst. 2015; 10:1-8.
  18. Takayama T, Ichihara T, Motomura N, Ogawa K. Determination of Energy Window Width and Position for Scintigraphic Imaging Using Different Energy Resolution Detection with the 'Triple Energy Winddw (TEW) Scatter Compensation Method. IEEE. 1998; 35(2):51-9.
  19. Changizi V, Takavar A, Babakhani A, Sohrabi M. Scatter correction for heart SPECT images using TEW method. J Appl Clin Med Phys. 2008; 9(3):2767.
  20. Dewaraja Yk, Ljungberg M, Fessler JA. 3-D Monte Carlo-based scatter compensation in quantitative I-131 SPECT reconstruction. IEEE Trans Nucl Sci. 2006; 53:181-8.
  21. Dewaraja Yk. Quantitative 131I SPECT with triple energy window Compton scatter correction. IEEE Trans. Nucl. Sci. 1998; 45:3109.
  22. Ljungberg M, King MA, Hademenos J, Strand SE. Comparison of four scatter correction methods using Monte Carlo simulated source distribution. J Nucl Med. 1994; 35:143-51.
  23. Fujioka H, Inoue T, Ishimaru Y, Akamune A, Murase K, Tanada S, et al. Compton scatter correction using the triple energy window (TEW) method in conventional single photon emission computed tomography without TEW acquisition hardware. Kaku Igaku. 1997; 34(4):251-8.
  24. Ogawa K. Simulation study of triple-energy-window scatter correction in combined TI-201, Tc-99m SPECT. Ann of Nucl Med. 1994; 8(4):277-81.
  25. Takahashi A, Himuro K, Yamashita Y, Komiya I, Baba S, Sasaki M. Monte Carlo simulation of PET and SPECT imaging of 90Y. Med Phys. 2015; 42(4):1926-35.
  26. Noori-Asl M, Sadremomtaz A, Bitarafan-Rajabi A. Evaluation of three scatter correction methods based on estimation of photopeak scatter spectrum in SPECT imaging: a simulation study. Phys Med. 2014; 30(8):947-53.
  27. Ljungberg M. The SIMIND Monte Carlo program Home Page. Available from: https://www.msf.lu.se/forskning/the-simind-monte-carlo-program.
  28. Soares DCF, Oliveira MCD, Santos RGD, Andrade MS, Vilela JMC, Cardoso VN, et al. Liposomes radiolabeled with 159Gd-DTPA-BMA: Preparation, physicochemical characterization, release profile and in vitro cytotoxic evaluation. Eur J Pharm Sc. 2011;42: 462–9.
  29. Soares DCF, Oliveira MCD, Barros ALBD, Cardoso VN, Ramaldes GA. Liposomes radiolabeled with 159Gd: In vitro antitumoral activity, biodistribution study and scintigraphic image in Ehrlich tumor bearing mice. Eur J Pharm Sc. 2011. 43:290–6.
  30. Galvão I, Santos  RG, Neves MJ. Evaluation of Gd and Gd159 as new approaches for cancer treatment. Inter  Nucl Atl Conf. 2011;24-8.

31. Cipreste MF, Peres AM, Cotta AAC, AragonFH, Antunes ADM, Leal AS, et al. Synthesis and characterization of 159Gd-doped hydroxyapatite nanorods for bioapplications as theranostic systems. Mat Chem Phy. 2016; 181:301-11.