An Estimate of Radiation Dose to the Lens of the Eyes, Parotid Gland, and Thyroid Gland in Dental Panoramic Radiography

Document Type : Original Paper

Authors

1 Student Research Committee, Lorestan University of Medical Sciences, Khorramabad, Iran

2 Student Research Committee, Dezful University of Medical Sciences, Dezful, Iran

3 Department of Medical Physics, Lorestan University of Medical Sciences, Khorram Abad, Iran

Abstract

Introduction: Dental panoramic radiography (DPR) is one of the most frequent diagnostic X-ray procedures, the application of which is currently on a growing trend. During DPR, several radiosensitive tissues, such as the lens of the eyes, parotid gland, and thyroid gland, contribute to the radiation field, and it is necessary to monitor their received dose. The aim of this study was to evaluate the radiation dose to the lens of the eyes, parotid gland, and thyroid gland in patients undergoing DPR at Lorestan Province, Western Iran.
Material and Methods: This cross-sectional study was performed on 180 patients of both genders referred to DPR at two most crowded hospitals in Khorramabad, Iran, namely Tamin-e Ejtemaei (TE) and Shohada-ye Ashayer (SA) hospitals. The radiation dose measurements were carried out using LiF (Mg, Cu, P) thermoluminescent dosimeters (TLDs). To measure the absorbed dose received by the lens of the eyes, parotid gland, and thyroid gland in each patient, five sets of three TLDs, wrapped in a thin plastic bag, were positioned over each eyelid and the anatomical position of the parotid and thyroid glands. The TLDs were read within 24 h of exposure.
Results: The mean absorbed dose received by the lens of the eyes, parotid gland, and thyroid gland were obtained as 155, 160, and 72 µGy for the TE Hospital, respectively. These values were obtained as l24, 558, and 56 µGy, respectively, for the SA Hospital. The results revealed a statistically significant difference between the organs located outside and inside the primary beam in terms of the absorbed dose (P<0.001).
Conclusion: The absorbed dose received by the lens of the eyes and thyroid gland was generally lower than the values reported in similar studies. Nevertheless, the absorbed dose received by the parotid gland in the SA Hospital exceeded the recommended dose reference level of 400 µGy in DPR.

Keywords

Main Subjects


  1.  

    1. Subbulakshmi AC, Mohan N, Thiruneervannan R, Naveen S, Gokulraj S. Positioning errors in digital panoramic radiographs: A study. Journal of Orofacial Sciences. 2016;8(1):22.
    2. Yamanel K, Arhun N, Arikan S. Assessment of 3,031 panoramic radiographs in an urban Turkish population: a community survey. Oral Radiology. 2009;25(1):30-6.
    3. Chaparian A, Dehghanzade F. Evaluation of radiation-induced cancer risk to patients undergoing intra-oral and panoramic dental radiographies using experimental measurements and Monte Carlo calculations. Int J Radiat Res. 2017;15(2):197-205.
    4. Mortazavi S, Ghiassi-Nejad M, Bakhshi M, Jafari-Zadeh M, Kavousi A, Ahmadi J, et al. Entrance surface dose measurement on the thyroid gland in orthopantomography: The need for optimization. Iran J Radiat Res. 2004;2(1):21-6.
    5. Moudi E, Haghanifar S, Hadian H, Shabestani Monfared A, Deilam G, Bahemmat N. Assessment of radiation exposure of Head and Neck during panoramic radiography in different age groups. Science Journal of Clinical Medicine. 2014;3(1):10-4.
    6. Bahreyni Toossi MT, Akbari F. Diagnostic Reference Level Arising from Dental Panoramic Radiography. Iranian Journal of Medical Physics. 2012;9(3):161-7.
    7. Mashood EB, Deilam KH, Nika MS, Ali DG, Bashodi ME. Assessment of radiation exposure of eyes, parotid and thyroid gland during panoramic radiography. Int J Otolaryngol Ophthalmol. 2016;2(1):13-8.
    8. Jibiri NN, Adeleye B, Kolude B. Radiation dose to the thyroid, eyes and parotid glands of patients undergoing intra-oral radiographic procedures in a teaching hospital in Ibadan, Oyo state Nigeria. Int J Radiat Res. 2017;15(1):101-6.
    9. Mohan N, Ravikumar P, Madhumitha C. Genotoxic and cytotoxic effects following dental and panoramic radiography. Indian Journal of Oral Sciences. 2016;7(2):92.
    10. Holroyd J. Trends in dental radiography equipment and patient dose in the UK and republic of Ireland. 2013.
    11. Gholami M, Maziar A, Khosravi H, Ebrahimzadeh F, Mayahi S. Diagnostic reference levels (DRLs) for routine X-ray examinations in Lorestan province, Iran. INTERNATIONAL JOURNAL OF RADIATION RESEARCH. 2015;13(1):85-90.
    12. Toossi M, Malekzadeh M. Radiation dose to newborns in neonatal intensive care units. Iranian Journal of Radiology. 2012;9(3):145-9.
    13. ICRP (1996). International Commission on Radiological Protection. Rdiological Protection and Safety in Medicine. ICRP Publication 73. Annals of the ICRP 26, No. 2 (Pergamon Press, Oxford).
    14. supporting Guidance I. 2.Diagnostic reference levels in medical imaging: Review and additional advice. ICRP committee. 2002;3:1-2.
    15. BahreyniToossi MT, Malekzadeh M. Local Diagnostic Reference Levels for Common Pediatric X-Ray Examinations in Khorasan Razavi Province, Iran. Iranian Journal of Medical Physics. 2014;11(4):301-7.
    16. COR NI. National protocol for patient dose measurements in diagnostic radiology. Chilton, UK: NRPB. 1992.
    17. Joyce M, McEntee M, Brennan PC, O’Leary D. Reducing dose for digital cranial radiography: The increased source to the image-receptor distance approach. Journal of Medical Imaging and Radiation Sciences. 2013;44(4):180-7.
    18. Looe HK , Eenboom F, Chofor N, Pfaffenberger A, Steinhoff M, Ruhmann A, et al. Conversion coefficients for the estimation of effective doses in intraoral and panoramic dental radiology from dose-area product values. Radiat Prot Dosimetry. 2008;131(3):365-73.
    19. Talaeipour AR, Abbasi VM, Sakhdari S, Valaei N, Jafari zadeh M. Comparison of absorbed dose in target organs using conventional and digital panoramic radiography. J Res Dent Sci. 2013;10(2):83-9.
    20. White SC, Pharoah MJ. Oral radiology : principles and interpretation. 4th ed. New Delhi : Mosby, chapter 3. 2014.
    21. Martinez-Beneyto Y, Alcarlz- Baños M, Pérez-Lajarin L, Rushton V. Clinical justification of dental radiology in adult patients: A review of the literature. Med Oral Patol Oral Cir Bucal. 2007;12:244-51.
    22. Gonzalez L VE, Fernandez R. Reference doses in dental radiodiagnostic facilities. Br J Radiol. 2001;74(878):153-6.
    23. Gijbels F, Jacobs R, Bogaerts R, Debaveye D, Verlinden S, Sanderink G. Dosimetry of digital panoramic imaging. Part I: patient exposure. Dentomaxillofacial Radiology. 2005;34(3):145-9.