Utilization of Electronic Portal Imaging Device (EPID) For Setup Verification and Determination of Setup Margin in Head and Neck Radiation Therapy

Document Type : Original Paper

Authors

1 Department of Medical Physics, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran

2 Department of medical physics, Faulty of medicine, Mashhad University of Medical Sciences, Mashhad, Iran

3 Nuclear Medicine Research Center, Mashhad University of Medical Sciences, Mashhad, Iran

Abstract

Introduction: Radiation therapy involves a multistep procedure; therefore, the error in patient set up is an inherent part of the treatment. Main purpose of this study was to determine the clinical target volume (CTV) to planning target volume (PTV) in head and neck cancer patients.
Material and Methods: A total of 15 patients who had daily portal images during the treatment courses were randomly selected in the present study. Systematic (Σ) and random (σ) errors were evaluated in three directions. The Isogray treatment planning system and Elekta linear accelerator were used in this study. Moreover, we had used MOSIAQ software as arecord and Verify system. Setup margins were calculated using three published margin recipes, including the International Commission on Radiation Units and Measurements (ICRU) report 62, as well as Stroom’s and van Herk’s formulae.
Results: Average magnitude of the translational errors was reported between 0.7 and 10 mm. The systematic and random errors for head and neck cancer patients were 3.55 (2.58-4.52) and 1.83 (1.56-2.10) mm, respectively. According to the ICRU report 62, as well as Stoorm’s and van Herk’s formulas, the required margins to cover the target were obtained within the ranges of 3.1-4.9, 6.4-10.5, and 7.7-12.7 mm, respectively.
Conclusion: According to the results of the present study, 6.5-10.5 mm extension in CTV to PTV margin can ensure that 90% of the head and neck cancer patients will receive a minimum cumulative CTV dose higher than or equal to 95% of the prescribed dose.

Keywords

Main Subjects


  1. Strbac B, Jokic VS. Evaluation of set-up errors in head and neck radiotherapy using electronic portal imaging. Physica Medica. 2013;29(5):531-6.
  2. Vejdani-Noghreiyan A, Ebrahimi-Khankook A. Developing the Adult Male ICRP Phantom and Evaluation the Absorbed Dose Received By Critical Organs in Head and Neck Region during the Radiotherapy of Eye Cancer. Iranian Journal of Medical Physics. 2016;13(3):211-7.
  3. Yin F-F, Rubin P, Schell MC, Wynn R, Raubertas RF, Uschold G, et al. An observer study for direct comparison of clinical efficacy of electronic to film portal images. International Journal of Radiation Oncology* Biology* Physics. 1996;35(5):985-91.
  4. Marks J, Haus A, Sutton H, Griem M. The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields. Cancer. 1976;37(6):2755-61.
  5. Shahedi F, Mommen nezhad M, Akbari H, Jamali M, Nasseri s. Mega Voltage Cone Beam Computed Tomography (MV- CBCT) using a Standard Medical Linear Accelerator and EPID: A feasibility study. Iranian Journal of Medical Physics. 2018;15(Special Issue-12th. Iranian Congress of Medical Physics):149.
  6. Shahedi F, Momennezhad M, Nasseri s, Gholamhosseinian H. Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy. Iranian Journal of Medical Physics. 2018;15(4):295-303.
  7. Burnet NG, Thomas SJ, Burton KE, Jefferies SJ. Defining the tumour and target volumes for radiotherapy. Cancer Imaging. 2004;4(2):153-61.
  8. Gildersleve J, Dearnaley D, Evans P, Swindell W. Reproducibility of patient positioning during routine radiotherapy, as assessed by an integrated megavoltage imaging system. Radiotherapy and Oncology. 1995;35(2):151-60.
  9. El-Gayed AA, Bel A, Vijlbrief R, Bartelink H, Lebesque JV. Time trend of patient setup deviations during pelvic irradiation using electronic portal imaging. Radiotherapy and Oncology. 1993;26(2):162-71.
  10. Brady LW, Perez CA, Wazer DE. Perez & Brady's principles and practice of radiation oncology: Lippincott Williams & Wilkins; 2013.
  11. Wambersie A, Landgerg T. ICRU Report 62: Prescribing, Recording and Reporting Photon Beam Therapy. ICRU Publ Bethesda MD. 1999.
  12. Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report. Radiotherapy and oncology. 2002;64(1):75-83.
  13. van Herk M, Remeijer P, Rasch C, Lebesque JV. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. International Journal of Radiation Oncology* Biology* Physics. 2000;47(4):1121-35.
  14. Gupta T, Chopra S, Kadam A, Agarwal JP, Devi PR, Ghosh-Laskar S, et al. Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device. Radiat Oncol. 2007;2(44):1-8.
  15. Van Herk M. Errors and margins in radiotherapy. InSeminars in radiation oncology 2004 Jan 1 (Vol. 14, No. 1, pp. 52-64). WB Saunders.
  16. Yan D, Ziaja E, Jaffray D, Wong J, Brabbins D, Vicini F, et al. The use of adaptive radiation therapy to reduce setup error: a prospective clinical study. International Journal of Radiation Oncology* Biology* Physics. 1998;41(3):715-20.
  17. Suzuki M, Nishimura Y, Nakamatsu K, Okumura M, Hashiba H, Koike R, et al. Analysis of interfractional set-up errors and intrafractional organ motions during IMRT for head and neck tumors to define an appropriate planning target volume (PTV)-and planning organs at risk volume (PRV)-margins. Radiotherapy and oncology. 2006;78(3):283-
  18. Hess CF, Kortmann R-D, Jany R, Hamberger A, Bamberg M. Accuracy of field alignment in radiotherapy of head and neck cancer utilizing individualized face mask immobilization: a retrospective analysis of clinical practice. Radiotherapy and Oncology. 1995;34(1):69-72.
  19. Bentel GC, Marks LB, Hendren K, Brizel DM. Comparison of two head and neck immobilization systems. International Journal of Radiation Oncology* Biology* Physics. 1997;38(4):867-73.
  20. Gilbeau L, Octave-Prignot M, Loncol T, Renard L, Scalliet P, Grégoire V. Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors. Radiotherapy and Oncology. 2001;58(2):155-62.
  21. de Boer HC, de Koste JRvS, Creutzberg CL, Visser AG, Levendag PC, Heijmen BJ. Electronic portal image assisted reduction of systematic set-up errors in head and neck irradiation. Radiotherapy and oncology. 2001;61(3):299-308.
  22. Humphreys M, Urbano MTG, Mubata C, Miles E, Harrington KJ, Bidmead M, et al. Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging. Radiotherapy and oncology. 2005;77(1):39-44.
  23. Zhang L, Garden AS, Lo J, Ang KK, Ahamad A, Morrison WH, et al. Multiple regions-of-interest analysis of setup uncertainties for head-and-neck cancer radiotherapy. International Journal of Radiation Oncology* Biology* Physics. 2006;64(5):1559-69.
  24. Babu N, Singh B, Namrata S, Mohanti B, Ravichandran R, Ghamrawy K. Development of patient support devices for execution of clinical radiotherapy for cancer patients: a preliminary report. Journal of Medical Physics/Association of Medical Physicists of India. 2006;31(4):255.
  25. Khosa R, Nangia S, Chufal KS, Ghosh D, Kaul R, Sharma L. Daily online localization using implanted fiducial markers and its impact on planning target volume for carcinoma prostate. Journal of cancer research and therapeutics. 2010;6(2):172.
  26. Osei E, Jiang R, Barnett R, Fleming K, Panjwani D. Evaluation of daily online set-up errors and organ displacement uncertainty during conformal radiation treatment of the prostate. The British journal of radiology. 2009;82(973):49-61.