Application of Patient-Customized Cast Type M3 Wax Bolus using a 3D printing for Photon Beam Radiation Therapy in Patients with Scalp Malignant Tumor

Document Type : Original Paper

Authors

1 Department of Radiation Oncology, Uijeongbu Eulji Medical Center, Eulji University 170, Uijeongbu-si, Gyeonggi-do, Korea

2 Department of Radiation Oncology, KonYang University Hospital, 158, Gwanjeodong-ro, Seo-gu, Daejeon, Korea

3 Department of Radiologic Technology, Dongnam Health University, 50, Cheoncheon-ro, Jangan-Gu, Suwon-si, Gyeonggi-do, Korea

4 Department of Radiological Science, Gachon University Medical Campus, Hambangmoe-ro, Yeonsu-gu, Incheon, Korea, 21936

Abstract

Introduction: We investigated the usefulness of patient-customized cast type M3 wax bolus (MWB) in radiation therapy in scalp malignant tumor patients by 3D conformal radiation therapy (3D CRT) and intensity-modulated radiation therapy (IMRT).
Material and Methods: A helmet-type polylactic acid (PLA) hollow model was fabricated using a 3D Printing, and the molten MWB was poured into the mold and allowed to harden. Subsequently, a solid MWB head cast was obtained by removing the PLA. The radiation volume was verified using a metal oxide semiconductor field-effect transistor (MOSFET) dosimeter and EBT3 film.
Results: Radiation dose verification was performed at the anterior, right, and left angles of planning tumor volume. The error rate demonstrated a maximum value of 5.5% and an average of 3.3% using the MOSFET dosimeter, and a maximum value of 7.0% and an average of 5.4% applying the EBT3 film. The homogeneity indices of the treatment plans were obtained as 0.09 and 0.12 using 3D CRT and IMRT, respectively. Moreover, the conformity number of the treatment plans was reported as 0.79 using 3D CRT and 0.81 applying IMRT.
Conclusion: The density of the MWB head cast was 1.05 g/cm3 which is closer to that of the equivalent tissue than the existing helmet type bolus material. In addition, it reduces the processing time and associated pain during custom manufacturing and has little air gaps. Therefore, it can be considered an effective method for the treatment of patients with scalp malignant tumors.

Keywords

Main Subjects


  1.  

    1. Wojcicka JB, Lasher DE, McAfee SS, Fortier GA. Dosimetric comparison of three different treatment techniques in extensive scalp lesion irradiation. Radiother Oncol. 2009 ;91(2):255-60.
    2. Caivano R, Fiorentino A, Pedicini P, Califano G, Fusco V. A radiotherapy technique for palliative total scalp irradiation. Ann Palliat Med. 2015;4(1):35-8.
    3. Majithia L, Rong Y, Siddiqui F, Hattie T, Gupta N, Weldon M, et al. Treating cutaneous T-cell lymphoma with highly irregular surfaces with photon irradiation using rice as tissue compensator. Front Oncol. 2015;5:49. DOI: 10.3389/fonc.2015.00049
    4. Sponseller P, Parvathaneni U. A case study of radiotherapy planning for intensity modulation radiation therapy for the whole scalp with matching electron treatment. Med Dosim. 2014;39(1):122-4.
    5. Vyas V, Palmer L, Mudge R, Jiang R, Fleck A, Schaly B, et al. On bolus for megavoltage photon and electron radiation therapy. Med Disum. 2013;38(3):268-73.
    6. Song JH, Jung JY, Park HW, Lee GW, Chae SM, Kay CS, et al. Dosimetric comparison of three different treatment modalities for total scalp irradiation: the conventional lateral photon-electron technique, helical tomotherapy, and volumetric-modulated arc therapy. J Radiat Res. 2014;56(4):717-26.
    7. Zhang RR, Feygelman V, Harris ER, Rao N, Moros EG, Zhang GG. Is wax equivalent to tissue in electron conformal therapy planning? A Monte Carlo study of material approximation introduced dose difference. J Appl Clin Med Phys. 2013; 14(1):92-101.
    8. Oh SA, Lee CM, Lee MW, Lee YS, Lee GH, Kim SH, et al. Fabrication of a Patient-Customized Helmet with a Three-Dimensional Printer for Radiation Therapy of Scalp. Prog Med Phys. 2017;28(3):100-5.
    9. White DR. Tissue substitutes in experimental radiation physics. Med Phys. 1978;5(6):467-79.
    10. Salimi M, Abi KST, Nedaie HA, Hassani H, Gharaati H, Samei M, et al. Assessment and Comparison of Homogeneity and Conformity Indexes in Step-and-Shoot and Compensator-Based Intensity Modulated Radiation Therapy (IMRT) and Three-Dimensional Conformal Radiation Therapy (3D CRT) in Prostate Cancer. J Med Signals Sens. 2017;7(2):102-7
    11. Feuvret L, Noel G, Mazeron JJ, Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys. 2006;64(2):333-42.
    12. IAEA TRS-398 (V.12). Absorbed Dose Determination in External Beam Radiotherapy:An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water. 2006.
    13. Dreindl R, Georg D, Stock M. Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films. Z Med Phys. 2014 ;24(2):153-63.
    14. Rudat V, Nour A, Alaradi AA, Mohamed A, Altuwaijri S. In vivo surface dose measurement using GafChromic film dosimetry in breast cancer radiotherapy: comparison of 7-field IMRT, tangential IMRT and tangential 3D-CRT. Radiat Oncol. 2014;15(9):156.
    15. Consorti R, Petrucci A, Fortunato F, Soriani A, Marzi S, Iaccarino G, et al. In vivo dosimetry with MOSFETs: dosimetric characterization and first clinical results in intraoperative radiotherapy. Int J Radiat Oncol Biol Phys. 2005;63(3):952-60.
    16. Choi JY, Won YJ, Park JY, Kim JW, Moon BK, Yoon HG, et al. Development of a Thermoplastic Oral Compensator for Improving Dose Uniformity in Radiation Therapy for Head and Neck Cancer. Prog Med Phys. 2012;23(4):269-78.
    17. Samant RS, Fox GW, Gerig LH, Montgomery LA. Allan DS. Total scalp radiation using image-guided IMRT for progressive cutaneous T cell lymphoma. Br J Radiol. 2009;82(978):122-5.
    18. Chen YJ, Liu A, Han C, Tsai PT, Schultheiss TE, Pezner RD, et al. Helical Tomotherapy for Radiotherapy in Esophageal Cancer: A Preferred Plan With Better Conformal Target Coverage and More Homogeneous Dose Distribution. Med Dosim. 2007;32(3):166-71.
    19. Fujimoto K, Shiinoki T, Yuasa Y, Hanazawa H, Shibuya K. Efficacy of patient-specific bolus created using three-dimensional printing technique in photon radiotherapy. Phys Med. 2017;38:1-9.
    20. Varadhan R, Miller J, Garrity B, Weber M. In vivo prostate IMRT dosimetry with MOSFET detectors using brass buildup caps. J Appl Clin Med Phys. 2006;7(4):22-32.

     

Volume 17, Issue 6
November and December 2020
Pages 428-434
  • Receive Date: 03 May 2019
  • Revise Date: 20 September 2019
  • Accept Date: 26 September 2019