Introduction: This study conducted a clinical accuracy analysis of Synchrony Respiratory Tracking System across different CyberKnife generations (G3, G4, VSI, and M6) to provide comprehensive comparing across different generations. Additionally, appropriate regression models were developed to explain the behavior of standard deviation (SD) of the errors (correlation and prediction) and tumor motion in each region. Material and Methods: The clinical log data was analyzed to assess the correlation and prediction errors. A retrospective analysis was conducted on 46 patients with thoracic or abdominal cancers treated using the CyberKnife G3. Furthermore, the F-test, P-value, and R2 analysis were applied to model the SD of correlation and prediction errors as a function of tumor displacements across five lung regions and two abdomen regions. Results: Using a systematic approach, linear, quadratic, cubic, or piecewise regression models were proposed for SD of the errors across tumor displacement. The estimated radial Synchrony error (mean±SD) for the CyberKnife G3, G4, VSI, and M6 was 2.60±0.93 mm, 2.00±0.60 mm, 1.79±1.16 mm, and 0.66±0.23 mm, respectively. Conclusion: The results indicate that correlation error remains predominant in both lung and abdominal regions across all CyberKnife generations. However, prediction error has been significantly reduced in the G4, VSI, and M6 systems. This improvement is attributed to the newer generations incorporating a combination of historical pattern-matching filters and least-mean-square filters. Error modeling based on tumor motion in different regions reveals a linear relationship between errors and target amplitude in the lung region, while the liver and pancreas regions exhibit non-linear relationships.
Chang JY, Senan S, Paul MA, Mehran RJ, Louie AV, Balter P, et al. Stereotactic ablative radiotherapy versus lobectomy for operable stage I non-small-cell lung cancer: a pooled analysis of two randomised trials. The Lancet Oncology. 2015;16(6):630-7.
Timmerman R, Paulus R, Galvin J, Michalski J, Straube W, Bradley J, et al. Stereotactic body radiation therapy for inoperable early stage lung cancer. Jama. 2010;303(11):1070-6.
Høyer M, Swaminath A, Bydder S, Lock M, Romero AM, Kavanagh B, et al. Radiotherapy for liver metastases: a review of evidence. International Journal of Radiation Oncology* Biology* Physics. 2012;82(3):1047-57.
Miandoab PS, Torshabi AE, Nankali S. Investigation of the optimum location of external markers for patient setup accuracy enhancement at external beam radiotherapy. Journal of Applied Clinical Medical Physics. 2016;17(6):32-43.
Rosenzweig KE, Hanley J, Mah D, Mageras G, Hunt M, Toner S, et al. The deep inspiration breath-hold technique in the treatment of inoperable non–small-cell lung cancer. International Journal of Radiation Oncology* Biology* Physics. 2000;48(1):81-7.
Underberg RW, Lagerwaard FJ, Slotman BJ, Cuijpers JP, Senan S. Benefit of respiration-gated stereotactic radiotherapy for stage I lung cancer: an analysis of 4DCT datasets. International Journal of Radiation Oncology* Biology* Physics. 2005;62(2):554-60.
Schweikard A, Shiomi H, Adler J. Respiration tracking in radiosurgery. Medical physics. 2004;31(10):2738-41.
Hoogeman M, Prévost J-B, Nuyttens J, Pöll J, Levendag P, Heijmen B. Clinical accuracy of the respiratory tumor tracking system of the cyberknife: assessment by analysis of log files. International Journal of Radiation Oncology* Biology* Physics. 2009;74(1):297-303.
Keall PJ, Mageras GS, Balter JM, Emery RS, Forster KM, Jiang SB, et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76 a. Medical Physics. 2006;33(10):3874-900.
Kilby W, Naylor M, Dooley JR, Maurer Jr CR, Sayeh S. A technical overview of the CyberKnife system. Handbook of Robotic and Image-Guided Surgery. 2020:15-38.
Kilby W, Dooley J, Kuduvalli G, Sayeh S, Maurer Jr C. The CyberKnife® robotic radiosurgery system in 2010. Technology in Cancer Research & Treatment. 2010;9(5):433-52.
Coste‐Manière È, Olender D, Kilby W, Schulz R. Robotic whole body stereotactic radiosurgery: clinical advantages of the CyberKnife® integrated system. The International Journal of Medical Robotics and Computer Assisted Surgery. 2005;1(2):28-39.
Sayeh S, Wang J, Main WT, Kilby W, Maurer CR. Respiratory motion tracking for robotic radiosurgery. Treating tumors that move with respiration: Springer; 2007. p. 15-29.
Pepin EW, Wu H, Zhang Y, Lord B. Correlation and prediction uncertainties in the cyberknife synchrony respiratory tracking system. Medical Physics. 2011;38(7):4036-44.
Winter JD, Wong R, Swaminath A, Chow T. Accuracy of robotic radiosurgical liver treatment throughout the respiratory cycle. International Journal of Radiation Oncology* Biology* Physics. 2015;93(4):916-24.
Nakayama M, Nishimura H, Mayahara H, Nakamura M, Uehara K, Tsudou S, et al. Clinical log data analysis for assessing the accuracy of the CyberKnife fiducial-free lung tumor tracking system. Practical radiation oncology. 2018;8(2):e63-e70.
Bibault J-E, Prevost B, Dansin E, Mirabel X, Lacornerie T, Lartigau E. Image-guided robotic stereotactic radiation therapy with fiducial-free tumor tracking for lung cancer. Radiation Oncology. 2012;7(1):1-7.
Chan M, Grehn M, Cremers F, Siebert F-A, Wurster S, Huttenlocher S, et al. Dosimetric implications of residual tracking errors during robotic SBRT of liver metastases. International Journal of Radiation Oncology* Biology* Physics. 2017;97(4):839-48.
Floriano A, García R, Moreno R, Sánchez‐Reyes A. Retrospective evaluation of CTV to PTV margins using CyberKnife in patients with thoracic tumors. Journal of Applied Clinical Medical Physics. 2014;15(6):59-72.
Liu M, Cygler JE, Vandervoort E. Geometrical tracking accuracy and appropriate PTV margins for robotic radiosurgery of liver lesions by SBRT. Acta Oncologica. 2019;58(6):906-15.
Yang Z-Y, Chang Y, Liu H-Y, Liu G, Li Q. Target margin design for real-time lung tumor tracking stereotactic body radiation therapy using CyberKnife Xsight Lung Tracking System. Scientific reports. 2017;7(1):1-11.
Samadi Miandoab P, Saramad S, Setayeshi S. Target margin design through analyzing a large cohort of clinical log data in the cyberknife system. Journal of Applied Clinical Medical Physics. 2022;23(3):e13476.
Samadi Miandoab P, Saramad S, Setayeshi S, Blanck O. A retrospective multi‐center feasibility study of a new PTV margin estimation approach for moving targets using CyberKnife log files. Journal of Applied Clinical Medical Physics. 2023;24(7):e13975.
Suh Y, Dieterich S, Cho B, Keall PJ. An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients. Physics in Medicine and Biology. 2008;53(13):3623.
Cameron AC, Windmeijer FA. An R-squared measure of goodness of fit for some common nonlinear regression models. Journal of econometrics. 1997;77(2):329-42.
Yang B, Chiu TL, Law WK, Geng H, Lam WW, Leung TM, et al. Performance evaluation of the CyberKnife system in real-time target tracking during beam delivery using a moving phantom coupled with two-dimensional detector array. Radiological physics and technology. 2019;12(1):86-95.
Murphy MJ, Dieterich S. Comparative performance of linear and nonlinear neural networks to predict irregular breathing. Physics in Medicine and Biology. 2006;51(22):5903.
Malinowski KT, McAvoy TJ, George R, Dieterich S, D'Souza WD. Mitigating errors in external respiratory surrogate-based models of tumor position. International Journal of Radiation Oncology* Biology* Physics. 2012;82(5):e709-e16.
Samadi Miandoab, P. , Setayeshi, S. and Saramad, S. (2025). Clinical Accuracy Analysis and Error Modeling in the CyberKnife System; A Comparative Study. Iranian Journal of Medical Physics, 22(1), 36-43. doi: 10.22038/ijmp.2024.80469.2411
MLA
Samadi Miandoab, P. , , Setayeshi, S. , and Saramad, S. . "Clinical Accuracy Analysis and Error Modeling in the CyberKnife System; A Comparative Study", Iranian Journal of Medical Physics, 22, 1, 2025, 36-43. doi: 10.22038/ijmp.2024.80469.2411
HARVARD
Samadi Miandoab, P., Setayeshi, S., Saramad, S. (2025). 'Clinical Accuracy Analysis and Error Modeling in the CyberKnife System; A Comparative Study', Iranian Journal of Medical Physics, 22(1), pp. 36-43. doi: 10.22038/ijmp.2024.80469.2411
CHICAGO
P. Samadi Miandoab , S. Setayeshi and S. Saramad, "Clinical Accuracy Analysis and Error Modeling in the CyberKnife System; A Comparative Study," Iranian Journal of Medical Physics, 22 1 (2025): 36-43, doi: 10.22038/ijmp.2024.80469.2411
VANCOUVER
Samadi Miandoab, P., Setayeshi, S., Saramad, S. Clinical Accuracy Analysis and Error Modeling in the CyberKnife System; A Comparative Study. Iranian Journal of Medical Physics, 2025; 22(1): 36-43. doi: 10.22038/ijmp.2024.80469.2411