Impact of Patient Arm Positioning on Ddose Distribution in Target and Organs at Risk in Right Bbreast Cancer Radiotherapy

Document Type : Original Paper

Authors
1 Biomedical Engineering and Medical Physics Department, School of Medicine, Shahid Beheshti University of Medical Sciences,Tehran, Iran
2 Medical Physics Department, Radiation Oncology Center, Vali Asr Hospital, Qom, Iran
3 Department of Radiation Oncology, Winship Cancer Institute, Emory University, Atlanta, Georgia, USA
10.22038/ijmp.2025.80785.2416
Abstract
Introduction: Breast cancer (BC) ranks as the second most frequently diagnosed cancer among women globally. The aim of this study was to evaluate the effect of the patient’s arm positioning on dose distribution in the organs at risk (OARs) and target in BC radiotherapy for mastectomy patients.
Material and Methods: Computed tomography (CT) scans of images of 30 right BC patients were collected. The patients were divided into two groups, each with 15 patients. The first group was with right arm-up positioning, and the second was with both arms-up positioning. The Monaco treatment planning system (TPS) was used for treatment planning. Radiotherapy included a prescribed dose of 50 Gy/25 fractions. For the planning target volume (PTV), the minimum dose (Dmin), mean dose (Dmean), maximum dose (Dmax), homogeneity index (HI), conformity index (CI), and tumor control probability (TCP) were calculated. For the OARs Dmin, Dmean, Dmax, and V20 Gy (%) for the lung and normal tissue complication probability (NTCP) were assessed. We compared the two groups based on PTV and OAR levels.
Results: There was no statistical difference between quantities for PTV and OARs for the lung, heart, and spinal cord, but a statistical difference was only found in Dmean and Dmax for the thyroid gland.
Conclusion: Although a statistically significant difference in thyroid dose was found, the absolute difference is small and within tolerance limits. Both positioning techniques provide comparable coverage of the target volume and sparing of the critical organs. With regard to patient comfort and ease of reproducibility during radiotherapy treatments, positioning with both arms up is generally favored, while positions with the right arm up may be an alternative if thyroid dose minimization is an issue.
Keywords
Subjects

1.        Duzenli C, Koulis T, Menna T, Carpentier E, Arora T, Coope R, et al. Reduction in doses to organs at risk and normal tissue during breast radiation therapy with a carbon-fiber adjustable reusable accessory. Pract Radiat Oncol. 2021;11(6):470-479.
2.        Zhang Q, Liu J, Ao N, Yu H, Peng Y, Ou L, et al. Secondary cancer risk after radiation therapy for breast cancer with different radiotherapy techniques. Sci Rep. 2020;10(1):1220-1226.
3.        Ma Z, Zhang W, Su Y, Liu P, Pan Y, Zhang G, et al. Optical surface management system for patient positioning in interfraction breast cancer radiotherapy. Biomed Res Int. 2018;9(1):64-71.
4.        Cheng CW, Das IJ, Stea B. The effect of the number of computed tomographic slices on dose distributions and evaluation of treatment planning systems for radiation therapy of intact breast. Int J Radiat Oncol Biol Phys. 1994;30(1):183-195.
5.        Mulliez T, Speleers B, Madani I, De Gersem W, Veldeman L, De Neve W. Whole breast radiotherapy in prone and supine position: Is there a place for multi-beam IMRT.  Radiat Oncol. 2013;8(1):151-158.
6.        Varga Z, Cserhati A, Rarosi F, Boda K, Gulyas G, Egyud Z, et al. Individualized positioning for maximum heart protection during breast irradiation. Acta Oncol. 2014;53(1):58-64.
7.        Chen K, Wei J, Ge C, Xia W, Shi Y, Wang H, et al. Application of auto-planning in radiotherapy for breast cancer after breast-conserving surgery. Sci Rep. 2020;10(1):10927-10932.
8.        Chiuyo J, Lugendo I, Yusuph S, Muhogora W. Dosimetric evaluation of intact breast radiotherapy in early stages breast cancer patients using field-in-field and 3D-CRT enhanced dynamic wedge techniques: Experience at a tertiary hospital in Tanzania. Iran J Med Phys 2023;20(6):333-340.
9.        Erraoudi M, Bouzekraoui Y, Asmi H, Bentayeb F. CTV to PTV margins based on CBCT method for prostate cancer of patients treated with VMAT technique. Iran J Med Phys 2023;20(5):266-272.
10.     Yao S, Zhang Y, Nie K, Liu B, Haffty BG, Ohri N, et al. Setup uncertainties and the optimal imaging schedule in the prone position whole breast radiotherapy.  Radiat Oncol. 2019;14(1):76-81.
11.     Speleers BA, Belosi FM, De Gersem WR, Deseyne PR, Paelinck LM, Bolsi A, et al. Comparison of supine or prone crawl photon or proton breast and regional lymph node radiation therapy including the internal mammary chain. Sci Rep. 2019;9(1):4755-4760.
12.     Kim H, Kim J. Evaluation of the anatomical parameters for normal tissue sparing in the prone position radiotherapy with small sized left breasts. Oncotarget. 2016;7(44):72211-72218.
13.     Errahmani MY, Locquet M, Broggio D, Spoor D, Jimenez G, Camilleri J, et al. Supraventricular cardiac conduction system exposure in breast cancer patients treated with radiotherapy and association with heart and cardiac chambers doses. Clin Transl Radiat Oncol. 2023;38(3):62-70.
14.     Errahmani MY, Locquet M, Spoor D, Jimenez G, Camilleri J, Bernier MO, et al. Association between cardiac radiation exposure and the risk of arrhythmia in breast cancer patients treated with radiotherapy: A case-control study. Front Oncol. 2022;12(1):89-93.
15.     Lamprecht B, Muscat E, Harding A, Howe K, Brown E, Barry T, et al. Comparison of whole breast dosimetry techniques - from 3dcrt to VMAT and the impact on heart and surrounding tissues. J Med Radiat Sci. 2022;69(1):98-107.
16.     Mankinen M, Virén T, Seppälä J, Hakkarainen H, Koivumäki T. Dosimetric effect of respiratory motion on planned dose in whole-breast volumetric modulated arc therapy using moderate and ultra-hypofractionation.  Radiat Oncol. 2022;17(1):46-51.
17.     Marrazzo L, Redapi L, Zani M, Calusi S, Meattini I, Arilli C, et al. A semi-automatic planning technique for whole breast irradiation with tangential IMRT fields. 2022;98(1):122-130.
18.     Niemierko A. Reporting and analyzing dose distributions: a concept of equivalent uniform dose. Med Phys. 1997;24:103-110.
19.     Emami B. Tolerance of normal tissue to therapeutic radiation. Rep. Radiother. Oncol. 2013;1(1):123-127.
20.     Petrova D, Smickovska S, Lazarevska E. Conformity index and homogeneity index of the postoperative whole breast radiotherapy. Open Access Maced. J. Med. Sci. 2017;5(6):736-739.
21.     Rudat V, Alaradi AA, Mohamed A, Ai-Yahya K, Altuwaijri S. Tangential beam IMRT versus tangential beam 3DCRT of the chest wall in postmastectomy breast cancer patients.  Radiat Oncol. 2011;6(1):26-30.
22.     Kazemzadeh A, Abedi I, Amouheidari A, Shirvany A. A radiobiological comparison of hypo-fractionation versus conventional fractionation for breast cancer 3D-conformal radiation therapy. Rep Pract Oncol Radiother. 2021;26(1):86-92.
23.     Nageeti T, Mahfouz M, Al Gaod M, Zatar R. Absorbed radiation dose in the thyroid gland following regional nodal irradiation for breast cancer. Int J Oncol Res. 2018;1(10):45-53
24.     Monaco Training Guide. Version 5.11.02. Available at: https://www.scribd.com/document/440748404/Monaco-5-10-Training-Guide. Accessed on: 1/1/2023.
Volume 23, Issue 2 - Serial Number 2
March and April 2026
Pages 103-110

  • Receive Date 26 June 2024
  • Revise Date 08 December 2025
  • Accept Date 22 July 2025