Assessment of Neutron Contamination in Small Fields of Elekta Synergy Linac in High Energy Photon Beams

Document Type : Original Paper

Authors
1 Department of Medical Physics, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran
2 Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences
3 Biomedical Engineering and Medical Physics Department, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran
4 Department of Radiation Oncology, Winship Cancer Institute, Emory University, Atlanta, GA, United States of America
10.22038/ijmp.2026.95945.2696
Abstract
Introduction: Since technique such as intensity-modulated radiation therapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT) use small fields (< 5 × 5 cm²), data on neutron contamination in such conditions remain limited. This study evaluated neutron dose equivalent (NDE) levels in small fields in 10 MV and 18 MV photon beam energies for Elekta synergy linac.
Material and Methods: Fast NDE were measured by using CR-39 detectors, calibrated with a ²⁴¹Am-Be neutron source. Measurements were conducted at five field sizes (1 × 1 to 5 × 5 cm²) in both 10 MV and 18 MV energies, at the isocenter that was aligned with the center of the field, 5 cm and 10 cm distances from the field edge, and at dmax in a SP34 solid water phantom (Scanditronix-Welhofer, Germany).
Results: The highest NDE (8.18 mSv/Gy) occurred at the isocenter (center of the field) for a 1 × 1 cm² field in 18 MV- five times higher than that for a 5 × 5 cm² field. NDE at dmax was lower by ~88% at the center of the field, and by ~90% at 10 cm from the field edge compared to the center of the field for a 1 × 1 cm2 field in 18 MV energy.
Conclusion: NDE increased with decreasing field sizes, at 10 MV and 18 MV energies. This effect should be considered in treatment planning for techniques that use small fields, due to its potential biological risks.
Keywords
Subjects

1.        Barton MB, Frommer M, Shafiq J. Role of radiotherapy in cancer control in low-income and middle-income countries. Lancet Oncol. 2006;7(7):584–95.
2.        De Ruysscher D, Niedermann G, Burnet NG, Siva S, Lee AW, Hegi-Johnson F. Radiotherapy toxicity. Nat Rev Dis Primers. 2019;5(1):13.
3.        Assalmi M, Diaf EY. The behavior of the contamination particles appears with the 18 MV photon beam delivered by Elekta Synergy MLCi2. Radiat Phys Chem. 2022;199:110384.
4.        Chen HH, Kuo MT. Improving radiotherapy in cancer treatment: Promises and challenges. Oncotarget. 2017;8(37):62742.
5.        Schaue D, McBride WH. Opportunities and challenges of radiotherapy for treating cancer. Nat Rev Clin Oncol. 2015;12(9):527–40.
6.        Moghaddam FF, Bakhshandeh M, Ghorbani M, Mofid B. Assessing the out-of-field dose calculation accuracy by Eclipse treatment planning system in sliding window IMRT of prostate cancer patients. Comput Biol Med. 2020;127:104052.
7.        Van Hoey O, Stolarczyk L, Lillhök J, Eliasson L, Mojzeszek N, Liszka M, et al. Simulation and experimental verification of ambient neutron doses in a pencil beam scanning proton therapy room as a function of treatment plan parameters. Front Oncol. 2022;12:903537.
8.        Khan FM. The physics of radiation therapy. 2nd ed. Lippincott Williams & Wilkins; 2009. p. 779–88.
9.        Rudd P, Prior D, Austin-Smith S. Neutron contamination of 10 MV X-rays: its relevance to treatment room door and maze design. Br J Radiol. 2007;80(954):469–75.
10.     Mesbahi A, Ghiasi H, Mahdavi SR. Photoneutron and capture gamma dose equivalent for different room and maze layouts in radiation therapy. Radiat Prot Dosimetry. 2010;140(3):242–9.
11.     Hashemi F, Hashemi-Malayeri B, Raisali G, Shokrani P, Sharafi A. A study of the photoneutron dose equivalent resulting from a Saturne 20 medical linac using Monte Carlo method. Nukleonika. 2007;52(1):39–43.
12.     NCRP. NCRP Report No. 93: Ionizing Radiation Exposure of the Population of the United States. Bethesda, MD: National Council on Radiation Protection and Measurements; 1993.
13.     Olsen KJ. Neutron contamination from medical electron accelerators (NCRP Report No. 79). Bethesda, MD: National Council on Radiation Protection and Measurements; 1986.
14.     Elmtalab S, Shanei A, Choopan Dastjerdi MH, Brkić H, Abedi I, Amouheidari A. Determination of the neutron contamination during brain radiotherapy using a moderated-boron trifluoride detector and the MCNP Monte Carlo code. Radiat Prot Dosimetry. 2022;198(3):129–38.
15.     Shojaei M, Rashedi S, Chakoli AN, Karimi AH, Geraily G. Measurement of neutron equivalent dose in the thyroid, chiasma, and lens for patients undergoing pelvic radiotherapy: A phantom study. Appl Radiat Isot. 2022;184:110188.
16.     Banaee N, Goodarzi K, Nedaie HA. Neutron contamination in radiotherapy processes: a review study. J Radiat Res. 2021;62(6):947–54.
17.     Vega-Carrillo H, Ortíz-Hernandez A, Hernandez-Davila V, Hernández-Almaraz B, Montalvo T. H*(10) and neutron spectra around linacs. J Radioanal Nucl Chem. 2010;283(2):537–40.
18.     ICRP. ICRP Publication 103: The 2007 Recommendations of the International Commission on Radiological Protection. Ann ICRP. 2007;37(2-4):1-332.
19.     Xu XG, Bednarz B, Paganetti H. A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction. Phys Med Biol. 2008;53(13):R193.
20.     Alfuraih A, Chin M, Spyrou N. Measurements of the photonuclear neutron yield of 15 MV medical linear accelerator. J Radioanal Nucl Chem. 2008;278:681–4.
21.     Ongaro C, Zanini A, Nastasi U, Ródenas J, Ottaviano G, Manfredotti C. Analysis of photoneutron spectra produced in medical accelerators. Phys Med Biol. 2000;45(12):L55.
22.     Newhauser WD, Durante M. Assessing the risk of second malignancies after modern radiotherapy. Nat Rev Cancer. 2011;11(6):438–48.
23.     Tajiki S, Nedaie HA, Rahmani F. A Monte Carlo study of neutron contamination in presence of circular cones during stereotactic radiotherapy with 18 MV photon beams. Biomed Phys Eng Express. 2020;6(3):035016.
24.     Hosseinzadeh E, Banaee N, Nedaie HA. Monte Carlo calculation of photo-neutron dose produced by circular cones at 18 MV photon beams. Rep Pract Oncol Radiother. 2018;23(1):39–46.
25.     Lazem BS, Hassan IJ, Rejah BK, Khalaf H, Mostafa MYA. Field Width Comparison of the Photon Beams 6 MV and 10 MV Coming Out of the Radiotherapy Linear Accelerator. Baghdad Sci J. 2025;22(7):2313–20.
26.     Gabrisch L, Cecchetto M, Delfs B, Looe HK, Budroweit J, Alía RG, et al. Quantification of neutron-induced single event upsets in a static random-access memory by clinical high-energy photon beam. IEEE Trans Nucl Sci. 2024;71(3):1-7.
27.     Assalmi M, Alshreef A, Diaf EY, Arectout A, Ade N, El Berhdadi EH. Quantitative Assessment of Photoneutron-induced Secondary Radiation Dose in Prostate Treatment Using an 18 MV Medical Linear Accelerator: A Monte Carlo Study. J Med Phys. 2024;49(4):673–86.
28.     Brkić H, Ivković A, Kasabašić M, Sovilj MP, Jurković S, Štimac D, et al. The influence of field size and off-axis distance on photoneutron spectra of the 18 MV Siemens Oncor linear accelerator beam. Radiat Meas. 2016;93:28–34.
29.     Ezzati AO, Studenski MT. Neutron damage induced in cardiovascular implantable electronic devices from a clinical 18 MV photon beam: a Monte Carlo study. Med Phys. 2017;44(11):5660–6.
30.     Jahangiri M, Hejazi P, Hashemi SM, Haghparast A, Hajizadeh B. The effect of field size and distance from the field center on neutron contamination in medical linear accelerator. Int J Adv Biol Biomed Res. 2015;3(1):97–104.
31.     Biltekin F, Yeginer M, Ozyigit G. Investigating in-field and out-of-field neutron contamination in high-energy medical linear accelerators based on the treatment factors of field size, depth, beam modifiers, and beam type. Phys Med. 2015;31(5):517–23.
32.     Jaffray DA, Gospodarowicz MK. Radiation therapy for cancer. In: Gelband H, Jha P, Sankaranarayanan R, Horton S, editors. Cancer: Disease Control Priorities, 3rd edition, Volume 3. Washington (DC): The International Bank for Reconstruction and Development/The World Bank; 2015.
33.     Negi A, Semwal A, Chandra S, Hariwal R, Sonkawade R, Kanjilal D, et al. Modifications induced by Li+3, Ni+9 and Au+9 ion beams to CR-39 polymer track detector. Radiat Meas. 2011;46(1):127–32.
34.     Nedaie HA, Darestani H, Banaee N, Shagholi N, Mohammadi K, Shahvar A, et al. Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations. J Med Phys. 2014;39(1):10–7.
35.     Sohrabi M, Hakimi A. Fast, epithermal and thermal photoneutron dosimetry in air and in tissue equivalent phantom for a high-energy X-ray medical accelerator. Z Med Phys. 2018;28(1):49–62.
36.     Bogard JS, Golzarri JI, Espinosa G. Dose calibration and track diameter distribution for 241Am-Be neutron source, using CR-39 nuclear track methodology. J Nucl Phys Mater Sci Radiat Appl. 2018;6(1):77–80.
37.     Khajetash B, Toossi MTB, Ghorbani M, Jahangiri M, Akbari F. Measurement of fast neutron contamination caused by the presence of wedge and block using CR-39 detector. J Cancer Res Ther. 2019;15(Suppl 1):S103–9.
38.     Fujibuchi T, Kodaira S, Sawaguchi F, Abe Y, Obara S, Yamaguchi M, et al. Measurement of the secondary neutron dose distribution from the LET spectrum of recoils using the CR-39 plastic nuclear track detector in 10 MV X-ray medical radiation fields. Nucl Instrum Methods Phys Res B. 2015;349:239–45.
39.     Mohammadi S, Behmadi M, Mohammadi A, Bahreyni Toossi MT. Thermal and fast neutron dose equivalent distribution measurement of 15-MV linear accelerator using CR-39 nuclear track detectors. Radiat Prot Dosimetry. 2020;188(4):503–7.
40.     Fleisher R, Price P, Walker R. Nuclear tracks in solids: Principles and applications. Berkeley; 1975.
41.     Israngkul-Na-Ayuthaya I, Suriyapee S, Pengvanich P. Evaluation of equivalent dose from neutrons and activation products from a 15-MV X-ray LINAC. J Radiat Res. 2015;56(6):919–26.
42.     Hashemi SM, Hashemi-Malayeri B, Raisali G, Shokrani P, Sharafi AA, Torkzadeh F. Measurement of photoneutron dose produced by wedge filters of a high energy linac using polycarbonate films. J Radiat Res. 2008;49(3):279–83.
43.     Kry SF, Howell RM, Salehpour M, Followill DS. Neutron spectra and dose equivalents calculated in tissue for high‐energy radiation therapy. Med Phys. 2009;36(4):1244–50.
44.     Li D, Deng X, Xue Y, Lou Z, Zhang Y, Guo W, et al. Neutron dose distribution in the treatment room for an accelerator in the flattening filter-free mode. Precis Radiat Oncol. 2017;1(1):13–9.
45.     Kim HS, Park YH, Koo BC, Kwon JW, Lee JS, Choi HS. Evaluation of the photoneutron field produced in a medical linear accelerator. Radiat Prot Dosimetry. 2007;123(3):323–8.
46.     Thekkedath SC, Raman RG, Musthafa M, Bakshi A, Pal R, Dawn S, et al. Study on the measurement of photo-neutron for 15 MV photon beam from medical linear accelerator under different irradiation geometries using passive detectors. J Cancer Res Ther. 2016;12(2):1060–4.
47.     Kumar V, Sonkawade RG, Dhaliwal AS. Optimization of CR-39 as neutron dosimeter. Indian J Pure Appl Phys. 2010;48(7):466-469.