The Effect of Breast Reconstruction Prosthesis on Photon Dose Distribution in Breast Cancer Radiotherapy

Document Type : Original Paper

Authors

1 Department of Biomedical Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran

2 Dept. of Medical Physics, Faculty of Medicine, Iran University of Medical Sciences

3 Associate Professor, Department of Radiotherapy and Oncology, Shaheed Beheshti University of Medical Sciences, Tehran, Iran

4 Department of Medical Physics, Tehran University of Medical Science, Tehran, Iran

Abstract

Introduction: Siliconeprosthetic implants are commonlyutilizedfor tissue replacement and breast augmentation after mastectomy. On the other hand, some patients require adjuvant radiotherapy in order to preventlocal-regional recurrence and increment ofthe overall survival. In case of recurrence, the radiation oncologist might have to irradiate the prosthesis.The aim of this study was to evaluate the effect of silicone prosthesis on photon dose distribution in breast radiotherapy.
Materials and Methods: The experimental dosimetry was performed using theprosthetic breast phantom and the female-equivalent mathematical chest phantom. A Computerized Tomographybased treatment planning was performedusing a phantom and by CorePlan Treatment Planning System (TPS). For measuring the absorbed dose, thermoluminescent dosimeter(TLD) chips (GR-207A) were used. Multiple irradiations were completed for all the TLD positions, and the dose absorbed by the TLDs was read by a lighttelemetry (LTM) reader.
Results: Statistical comparisons were performed between the absorbed dosesassessed by the TLDs and the TPS calculations forthe same sites. Our initial resultsdemonstratedanacceptable agreement (P=0.064) between the treatment planning data and the measurements. The mean difference between the TPS and TLD resultswas 1.99%.The obtained findings showed that radiotherapy is compatible withsilicone gel prosthesis.
Conclusion: It could be concludedthat the siliconbreast prosthesis has no clinicallysignificant effectondistribution of a 6 MV photon beam for reconstructed breasts.

Keywords

Main Subjects


  1. 1.       Krishnan L, George F, Mansfield C and Krishnan E. Effect of silicon gel breast prosthesis on electron and photon dose distributions. Medical Physics. 1983;10:96-9.DOI: 10.1118/1.595279.

    1. KuskePR, Schuster R, Klein EE, Young L, Perez C ,Fineberg B. Radiotherapy and breast reconstruction: clinical results and dosimetry. Int. J. Radiation Oncology Biol. Phys. 1991 Jul 1;21(2):339-46.DOI: 10.3978/j.issn.2227-684X.2012.05.02.
    2. Klein EE, Kuske PR. Changes in photon dose distributions due to breast prostheses. Int J Radiat Oncol Biol Phys.1993;25:541-9.DOI: 10.1016/0360-3016(93)90078-A.
    3. Oshaghi M, Sadeghi M, Mahdavi SR, Shirazi AR. Dosimetry of MammoSite applicator: Comparison between Monte Carlo simulation, measurements, and treatment planning calculation. Journal of Cancer Research and Therapeutics.2013;9:224.DOI: 10.4103/0973-1482.113361.
    4. Scutt D, Lancaster GA, Manning JT. Breast asymmetry and predisposition to breast cancer. Breast Cancer Res.2006;8:(2): R14.DOI: 10.1186/bcr1388.
    5. International Commision on Radiological Protection. Report of the Task Group on Reference Man, ICRP publication 23. Oxford: Pergamon Press;1975.
    6. Banaee N and Nedaie H.A. Evaluating the effect of energy on calibration of thermoluminescent dosimeters 7-LiF:Mg,Cu,P (GR-207A). International Journal of Radiation Research.2013;11:51-4.
    7. Bahreyni Toossi MT, Rajab Bolookat E, Salek R, Layegh M. Dose measurements of parotid glands and spinal cord in conventional treatment of nasopharyngeal carcinoma using rando phantom and thermoluminescent dosimeters. Iranian Journal of Medical Physics. 2015;12(2):78-84.DOI: 10.22038/ijmp.2015.4769.
    8. Chu FC, Kaufmann TP, Dawson GA, Kim YS, Rajaratnam S, Hoffman LA. Radiation therapy of cancer in prosthetically augmented or reconstructed breasts. Radiology.1992;185:429-33.DOI: 10.1148/radiology.185.2.1410349.